        Photovoltaic Auto safety comprehensive tester

        Product Description:Name:Photovoltaic Auto safety comprehensive testerType:BND-CS9933EG-15Standards: IEC61730; IEC61215; IEC 61646Description of PV safety tester:Photovoltaic safety tester is used to test PV module, cable, connector, terminal box, …etc which can deal with high voltage and high sensitivity insulation test. There are two functions that are high voltage test and insulation resistance test in total.
        Description

        Product Description:

        Name:Photovoltaic Auto safety comprehensive tester

        Type:BND-CS9933EG-15

        Standards: IEC61730; IEC61215; IEC 61646


        Description of PV safety tester:

        Photovoltaic safety tester is used to test PV module, cable, connector, terminal box, …etc which can deal with high voltage and high sensitivity insulation test. There are two functions that are high voltage test and insulation resistance test in total.

        According to the IEC61730 standard(photovoltaic module safety qualification—part I: Requirements for construction which describes the fundamental construction requirements for photovoltaic modules in order to provide safe electrical and mechanical operation during their expected lifetime), the definition of PV module application is divided into A, B, C.

        The current measuring terminal can choose the grounding mode or the floating mode. Insulation voltage withstand test of photovoltaic module is based on the application of IEC61215/61646 and the maximum voltage of the system.

        Maximum test voltage for the application class A:2000V plus 4 times the maximum system voltage

        For the maximum voltage application category B:1000V system 2 times

        According to the calculation of a maximum voltage of 8000V, CS9923GS, CS9923G is based on this standard production

        According to the calculation of the maximum voltage B 6000V, CS9922G, CS9922GS is based on this standard production


        Performance characteristics:

        ◆With safety anti-electric wall function, it can automatically detect whether the input power is correct

        ◆For photovoltaic system panels, controllers, and battery packs

        ◆After the DC withstand voltage and insulation resistance tests are completed, the tester can quickly discharge within 0.2 seconds to protect the safety of the operator

        ◆With current lower limit alarm function, it can prevent the open circuit of the test line from causing misjudgment of the tested components during the test. The output voltage has voltage upper limit and voltage lower limit alarm functions

        ◆With keyboard lock function to prevent the operator from modifying the test parameters and ensure that the tested object is tested according to the set parameters

        ◆Linear power amplifier drives the output of sinusoidal voltage (current), and the crest factor is within the range of 1.3 to 1.5

        ◆Adopts 240*64 graphic LCD display and humanized interface design

        ◆With Chinese and English display interfaces, it can meet the different needs of different users.

        ◆Test parameters can be saved in advance: 30 files can be set, and each file can have 99 test steps

        ◆Optional RS232, RS485, USB interface, standard PLC interface

        pv-auto-safety-tester (1).jpg


        The Specification of the PV safety test system:

        Model

        BND-CS9933EG-15









        ACW


        The output voltage

        Range

        0.050kV~6.500kV

        Accuracy

        ±(1.5% reading + 5V)

        Resolution

        1V

        Maximum output power

        130VA(6.500kV/20mA)

        Maximum rated current

        20mA

        Lower current range

        0~20mA, 0=no lower limit

        Current range

        200uA, 2mA, 20mA

        Output waveform

        Sine wave

        Output waveform distortion

        ≤2% (no load or pure resistive load)

        Crest factor

        1.3~1.5

        Output signal type

        DDS+ amplifier

        Voltage rise time

        0.3s~999.9s 0=voltage rise time off

        Test time

        0.3s~999.9s 0=continuous test

        Voltage fall time

        0.3s~999.9s 0=voltage fall time off

        Interval time

        0.0s~999.9s 0=interval time off

        Output voltage mode

        N-mode, G-mode







        DCW






        The output voltage

        Range

        0.050kV~6.500kV

        Accuracy

        ±(1.5%Value +5V)

        Resolution

        1V

        Maximum output power

        65W(6.500kV/10mA)

        Maximum rated current

        10mA

        Current range

        2uA、20uA、200uA、2mA、10mA

        Ripple factor

        ≤5%

        Discharge time

        ≤200ms

        Maximum charge current

        10mA

        Voltage rise time

        0.3s~999.9s 0=voltage rise time off

        Test time

        0.3s~999.9s 0=continuous test

        Voltage drop time

        0.3s~999.9s 0=voltage drop time off

        Interval time

        0.0s~999.9s 0=interval time off

        Delay alarm time

        0.3s~999.9s 0=delay alarm time off

        Output voltage mode

        N mode, G mode


        IR



        The output voltage

        Range

        0.050kV~1.500kV

        Accuracy

        ±(1.5% reading +5V)

        Resolution

        1V

        Maximum upper limit setting value

        50.00GΩ




        Maximum and minimum setting values

        50.00GΩ

        Minimum setting values

        1MΩ

        Voltage rise time

        0.3~999.9s 0=voltage rise time off

        Test time

        0.3s~999.9s 0=continuous test

        Interval time

        0.0s~999.9s 0=interval time off

        Automatic gear switching

        Can be set to on or off

        Discharge time

        ≤200ms







        GR


        Output current

        Range

        3.00A~52.00A(DC)

        Accuracy

        ±(1.5% Value +0.2A)

        Resolution

        0.01A

        Maximum output power

        250W(50A/0.10Ω)

        Maximum output current

        50AW

        Resistance upper limit setting value

        40A x150mW

        Set current value

        Output waveform

        Sine wave

        Output waveform distortion

        ≤2% (no load or pure resistive load)

        Crest factor

        1.3~1.5

        Output signal type

        DDS+power amplifier

        Test time

        0.3s~999.9s 0=continuous test

        Interval time

        0.3s~999.9s 0=interval time off

        Voltmeter


        Range

        0.050kV~6.500kV

        0.050kV~6.500kV

        0.050kV~6.500kV

        0.050kV~6.500kV

        Accuracy

        ±(1.5% reading +5V)

        Resolution

        1V

        Display value

        RMS value



        Withstand voltage ammeter

        Measure range

        AC

        0 ~ 20mA

        DC

        0 ~ 10mA


        Resolution

        AC

        200uA: 0.1uA, 2mA: 1uA, 20mA: 10uA

        DC

        2uA: 0.001uA, 20uA: 0.01uA, 200uA: 0.1uA, 2mA: 1uA,

        10mA: 10uA

        Measurement accuracy

        ≥2mA: ±(1.5%+5 digits), <2mA: ±(2%+5 digits)

        Offset function

        The current of the test leads and accessories can be subtracted.

        Test mode

        GND mode: RETURN terminal connected to the chassis

        FLOAT mode: RETURN terminal not connected to the chassis


        Measurement range

        1MΩ~50000MΩ


        Resolution

        1MΩ~9.999MΩ:0.001MΩ,10MΩ~99.99MΩ:0.01MΩ,

        100MΩ~999.9MΩ:0.1MΩ,1000MΩ~9999MΩ:1MΩ

        10GΩ~50GΩ:10MΩ

        Insulation resistance meter



        Accuracy

        0.100kV~0.200kV:1MΩ~999MΩ ±5%,1000MΩ~3000MΩ ±10%

        0.201kV~0.499kV:1MΩ~999MΩ ±5%,1000MΩ~5000MΩ ±10%

        0.500kV~1.000kV:1MΩ~999MΩ ±5%,1000MΩ~9999MΩ ±10%

        10000MΩ~50000MΩ ±15%

        Ground current meter

        Measuring range

        3.00A~52.00A

        Resolution

        0.01A

        Accuracy

        ±(1.5% of reading + 0.2A)

        Display value

        RMS value

        Ground resistance meter

        Range

        0~510mΩ

        Accuracy

        When the current is less than 5.00A, it is ± (3% of reading + 3mΩ) When the current is ≥ 5.00A, it is ± (2% of reading + 2mΩ)

        Resolution

        0.1mΩ

        Measuring method

        Four-terminal method

        Timer

        Range

        0~999.9s

        Resolution

        0.1s

        Accuracy

        ±(0.1%+50ms)